1. After wearing protective glasses and gloves, press to close the biological indicator cap.
2. Extrusion of the glass ampoule containing the medium in the crushing hole
3. Pinch the end of the biological indicator cap, tap the bottom of the vial on the table, moisten the medium with the spore piece at the bottom of the vial, do not knock the vial on the machine.
4, the device will warm up for 30 minutes after the biological indicator into an appropriate culture / reading aperture
5. Close the auto-reader lid and wait for the minus or plus sign to appear.
6. When the hole cover is open, all the red, yellow, and green lights will flash for one second to prove that they are working.
Do not remove the biological indicator before the test result is released. If the biological indicator is accidentally taken out before the end of the culture process, the machine will issue an alarm. The corresponding LED screen of the culture hole will prompt, and the C2 warning code will appear at the same time. You must put the indicator back in 10 seconds or your results will be lost!
7. There are three corresponding indications for each culture/reading hole: (+)(*) and (-), the biological indicator is placed in the reading hole of the corresponding color, and the corresponding LED screen shows the remaining time indicating that the culture/reading hole is being jobs.
8. Once a positive result is found (sterilization failure), (+) lights up and an alarm sounds (if the sound feature is turned on).
9. After the specified incubation time is over, (-) will illuminate to indicate successful sterilization.
10. The biological indicator can be taken out from the reading hole after (+) or (-) is lit, and do not move or touch the indicator before the detection result is obtained.
11. Once the result is found to be positive (+) will be displayed immediately, the negative (-) result indicates that the entire culture process is required.