Introduction to common faults and maintenance methods of electrothermal constant temperature incubator

Related Searches: Electric Thermostat Incubator Electric Thermostat Thermostatic Incubator Electrothermal Incubator Shanghai Basil Constant Temperature Incubator

The electrothermal constant temperature incubator is suitable for bacterial culture, fermentation and constant temperature test in scientific research and industrial production departments such as medical and health, pharmaceutical industry, biochemistry and agricultural science.

Heated Incubators

Common faults and maintenance methods of electrothermal constant temperature incubator :

1, no power

Cause: The socket has no power

Repair method: replace the socket

Fault reason: the plug is not plugged in well

Repair method: re-plug

Fault reason: fuse open circuit

Repair method: replace the fuse

2, the temperature inside the box does not rise

Cause of failure: setting the temperature too low

Maintenance method: readjust the set temperature

Cause of failure: electric heater is broken

Maintenance method: new electric heater

Cause of failure: The temperature controller has failed

Repair method: replace the new temperature controller

3, set the temperature and the temperature error inside the box is large

Cause of failure: temperature sensor is bad

Repair method: replace the temperature sensor

Cause of failure: temperature controller thyristor is bad

Maintenance method: change the temperature controller thyristor

4, do not cool down or slow down slowly

Cause of failure: insufficient refrigerant in the refrigeration system

Maintenance method: increase refrigerant

Cause of failure: dirty or ice blockage in the refrigeration system piping

Maintenance method: cleaning the refrigeration system pipeline

Cause of failure: The solenoid valve that supplies liquid to the evaporator is damaged.

Repair method: replace the new solenoid valve

5, the suction pressure is too low

Cause of failure: insufficient refrigerant in the refrigeration system

Maintenance method: supplement refrigerant

Cause of failure: expansion valve is blocked or damaged

Maintenance method: dry the pipeline or replace the expansion valve

Cause of the malfunction: the filter is clogged

Repair method: Replace the filter

6, the compressor does not work

The cause of the failure: the insurance of the compressor burned out

Maintenance method: replacement of compressor insurance

Cause of failure: insufficient power supply voltage

Maintenance method: replace the power supply to increase the voltage

Cause of malfunction: Contactor that controls compressor start-up is damaged

Repair method: replace the compressor start contactor

Note: The above must be repaired by professional maintenance personnel. Personally, please do not disassemble the cabinet.

For product information related to the electrothermal incubator , please see:

Http://  

For more technical and sales needs, please contact the relevant service staff of Shanghai Bashu Industrial Co., Ltd.

In 2007, Shanghai Bayu ( ) Industrial Co., Ltd. established an offline supply system for laboratory instruments, MRO industrial products, etc. In 2009, it officially launched an online procurement operation platform.

Hydrophilic Introducer Sheath Kit

Hydrophilic Introducer Sheath,introducer sheath,Hydrophilic,arterial sheath introducer

Anesthesia Medical Co., Ltd. , https://www.jssinoanesthesias.com

Posted on