New version of GMP B+A clean area

The reading requirements of the B+A clean area B+A in the new GMP are explained as follows: Class A (high-risk operation area) must reach static level 100, dynamic level 100, and the background area of ​​the area is grade B. That is, the background area is required to reach static level 100 and dynamic level.

Key words : B+A clean area Cleanliness detector 2.83L 28.3L dust particle counter plankton sampler

B+A level explanation:

1. Class A (high-risk operating areas, such as filling areas, areas where plastic stoppers are placed and open packaging containers in direct contact with sterile preparations, and areas for aseptic assembly or joining operations) are to achieve a cleanliness level of : Static 100, dynamic 100.

2, B + A cleanliness requirements are explained as: A level (high-risk operating area) to achieve static 100, dynamic hundred, and the background area of ​​the area is B, that is, the background area is required to reach static 100 , dynamic million.

3. It is not correct to understand B+A as dynamic to reach 10,000 levels and static to reach level 100.

The reference materials are as follows:

Traditional air cleanliness grading standards: GB/T16292-1996 (Chinese standard)

Particle size, numerical cleanliness level

Dust zui large allowable / cubic meter

Microbial Zui large allowable number

≥0.5um

≥5um

Floating bacteria / cubic meters

Settling bacteria/dish

Level 100

3,500

0

5

1

10,000 levels

350,000

2,000

100

3

100,000 level

3,500,000

20,000

500

10

300,000

10,500,000

60,000

-

15

Air cleanliness grading standard: ISO14644-1 (international standard)

Air cleanliness level (N)

Large concentration limit of particles greater than or equal to the particle size of the target particle size (units per cubic meter of air particles)

0.1 um

0.2 um

0.3 um

0.5um

1um

5 um

ISO Class 1

10

2

 

 

 

 

ISO Class 2

100

twenty four

10

4

ISO Class 3

1 000

237

102

35

8

ISO Class 4

10 000

2 370

1 020

352

83

ISO Class 5

100 000

23 700

10 200

3 520

832

29

ISO Class 6

1 000 000

237 000

102 000

35 200

8 320

293

ISO Class 7

 

352 000

83 200

2 930

ISO Class 8

 

3 520 000

832 000

29 300

ISO Class 9

 

35 200 000

8 320 000

293 000

Note: Due to the uncertainty involved in the measurement process, it is required to use no more than three valid concentration numbers to determine the level.

Sterile drug appendix:

The standards for air suspension particles of the above various levels are as follows:

Cleanliness level

Suspended particles zui large allowable / cubic meter

 

Static

dynamic

 

≥0.5μm

≥5.0μm

≥0.5μm

≥5.0μm

Class A

3520

20

3520

20

Class B

3520

29

352000

2900

Class C

352000

 

2900

3520000

29000

Class D

3520000

29000

Not regulated

Not regulated

 

Compare and learn:

Cleanliness level

Suspended particles zui large allowable / cubic meter

 

Static

dynamic

 

≥0.5μm

≥5.0μm

≥0.5μm

≥5.0μm

Class A

The traditional 100 grade, considering the particle size ≥ 5.0μm as ISO 4.8

The traditional 100 grade, considering the particle size ≥ 5.0μm as ISO4.8

Class B

Traditional 100, ISO 5

Traditional million, ISO 7

Class C

Traditional million, ISO 7

Traditional 100,000, ISO 8

Class D

Traditional 100,000, ISO 8

Not regulated

Class B: Refers to the background area where high-risk operations such as aseptic preparation and filling are in Class A clean areas.

Clean room cleanliness testing experts and dust particles planktonic bacteria sampler factory in Suzhou long stay Purification Technology Co. reminder: select the new GMP particle counter in B + A 2010 class clean area the new GMP requirements for airborne particle counters: To confirm Class A The level of clean area, the sampling amount of each sampling point shall not be less than 1 cubic meter (28.3L/min), and the level of airborne particles in Class A clean area shall be ISO 4.8, with the suspended particles of ≥5.0μm as the limit standard. The cleanliness of other clean rooms should be counters of not less than 2.83 L/min.

Planktonic bacteria sampler pharmaceutical plant cleanliness level and monitoring should be dynamic monitoring microorganisms, microbial assessment aseptic conditions. The airborne microbial sampler is based on the particle impact principle and the isokinetic sampling theory. The sampled air with microorganisms is sprayed by a pump at a high speed and impinged on a culture dish containing the medium, and colonies are formed after being cultured to count.

Portable Sprinkler irrigation system

Portable Sprinkler Irrigation System mainly used for Farmland irrigation, fruit and grass irrigation,Sugarcane sprinkler irrigation, power plants precipitation, factory and port dust removal.

Portable Sprinkler irrigation system available with multiple nozzles, allowing watering and misting options, irrigation booms are adaptable to any crop. Depending on the boom selected, multiple zones can be irrigated with one boom, allowing the grower to effectively grow multiple crops in one house. All booms can be controlled independently or through computerized controls.


Portable Sprinkler Irrigation System,Portable Drip Irrigation,Portable Irrigation System,Portable Sprinkler Irrigation

JIANGSU SKYPLAN GREENHOUSE TECHNOLOGY CO.,LTD , https://www.greenhousehydroponic.com

Posted on